• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION ORG-9100A; MULTIPLE PATIENT RECEIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION ORG-9100A; MULTIPLE PATIENT RECEIVER Back to Search Results
Model Number ORG-9100A
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/19/2016
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer (bme) reported that his telemetry system is having whole house intermittent signal loss.The bme requested on-site assistance to help isolate the specific device.No patient harm was reported during the signal loss periods.The two legs that were affecting the system were disconnected to allow the system to run without issues.During the on-site service, no signal loss events occurred.The system appeared to be operating within manufacturer's specifications.
 
Event Description
The biomedical engineer (bme) reported that his telemetry system is having whole house intermittent signal loss.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORG-9100A
Type of Device
MULTIPLE PATIENT RECEIVER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key6966968
MDR Text Key90696485
Report Number8030229-2017-00359
Device Sequence Number1
Product Code DRG
UDI-Device Identifier04931921103883
UDI-Public04931921103883
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberORG-9100A
Device Catalogue NumberORG-9100A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA10/20/2017
Distributor Facility Aware Date09/19/2016
Device Age82 MO
Event Location Hospital
Date Report to Manufacturer10/20/2017
Date Manufacturer Received10/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/27/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-