Model Number 5586
Device Problem Difficult to Remove (1528)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/27/2017
Event Type  malfunction  
Manufacturer Narrative
The reported inability to loosen the right ventricular set screw was confirmed in the laboratory. Visual inspection identified calcified blood filing the setscrew inset. The calcified blood was limiting wrench insertion and preventing the wrench from engaging the setscrew inset walls which is necessary to untighten the setscrew.
 
Event Description
It was reported that the patient presented in hospital for normal pulse generator change out. During the procedure, the physician was unable to loosen the sets screw. Multiple attempts were made but were unsuccessful. The lead was cut and device was replaced. The patient condition was good after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFRONTIER II
Type of DeviceIMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key6967089
MDR Text Key109297479
Report Number2017865-2017-33432
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/24/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date09/30/2008
Device Model Number5586
Device Lot Number0002134471
Other Device ID Number05414734502948
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-