• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND VALIANT CAPTIVIA; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND VALIANT CAPTIVIA; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number VAMF4646C100TU
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/26/2017
Type of Reportable Event Malfunction
Event or Problem Description
A valiant stent graft system was implanted in a patient for the endovascular treatment of a thoracoabdominal aneurysm.It was reported that the physician successfully implanted the valiant stent graft along with superior mesenteric artery (sma) and left renal artery with the snorkel technique.The physician modelled the valiant stent graft with ballooning however the aortogram revealed a type ia endoleak.The physician elected to snorkel the celiac artery, lengthen the sma stent and the left renal artery stent and implanted a 2nd valiant stent graft higher in the descending thoracic aorta.All devices were placed successfully.Ballooning was completed however the aortogram revealed that the type ia endoleak was still present.Per the physician, the cause of the proximal type i endoleak was due to the patient¿s anatomy, thoracoabdominal aneurysm.No additional clinical sequelae were reported and the patient will be monitored by the physician.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT CAPTIVIA
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key6967941
Report Number2953200-2017-01611
Device Sequence Number19099748
Product Code MIH
UDI-Device Identifier00643169347328
UDI-Public00643169347328
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/26/2019
Device Model NumberVAMF4646C100TU
Device Catalogue NumberVAMF4646C100TU
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2017
Initial Report FDA Received Date10/23/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
-
-