| Model Number |
G23773 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/06/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Pma/510(k) #: k122917.The event is currently under investigation.A follow-up report will be submitted upon receipt of additional information or completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The distributor reported that during inspection of the coda balloon catheter it was observed that there was a hair inside the primary package.This product was not used on a patient.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The qc specifications, device history record for the lot number involved was reviewed.There was one non conformance noted for foreign matter in the packaging seal, which was reworked.A complaint search revealed this complaint to be the only reported complaint associated to the complaint lot number 7853508.Retraining was conducted, due to the hair inside the primary package which was observed during inspection at the distribution facility.Manufacturing related/foreign matter has been determined as the root cause.We will notify the appropriate personnel and continue to monitor for similar complaints.No adverse events on patient was noticed due to this occurrence.
|
| |
|