| Catalog Number |
628223S |
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
09/28/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device will not be returned.If additional information becomes available it will be provided on a supplemental report.Device discarded by hospital.
|
| |
|
Event or Problem Description
|
|
It was reported that the 4th locking screw would not lock on the superior lateral plate variax clavicle 3 hole plate after locking 3 screws distally.The surgeon tried several times but the screw would not lock.The screw was not inserted.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event that superior lateral plate variax clavicle 3 hole / right was alleged of 'no locking effect' could not be confirmed, since the device was not returned (implanted) for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that the 4th locking screw would not lock on the superior lateral plate variax clavicle 3 hole plate after locking 3 screws distally.The surgeon tried several times but the screw would not lock.The screw was not inserted.
|
| |
|