BOSTON SCIENTIFIC - MAPLE GROVE MAVERICK²¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
|
Back to Search Results |
|
| Model Number |
H7493892812250 |
| Medical Device Problem Code |
Kinked (1339)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/04/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Device evaluated by mfr.: returned product consisted of a maverick 2 balloon catheter.The balloon was loosely folded with fluid in the balloon and lumen.The hypotube, distal shaft, balloon, markerbands, and tip were microscopically and tactile inspected.Inspection revealed a burst in the balloon material that was 8 mm in length and numerous kinks in the hypotube.Inspection of the remainder of the device revealed no other damage or irregularities.The most probable root cause is operational context as device performance was limited due to anatomical/procedural factors.(b)(4).
|
| |
|
Event or Problem Description
|
|
Reportable based on analysis completed on 04-oct-2017.It was reported that shaft kink occurred.Vascular access was obtained via the femoral artery.The 24x3.0mm, de novo target lesion was located in the none tortuous and severely calcified left circumflex (lcx) artery.A 2.50mm x 12mm maverick²¿ balloon catheter advanced for dilatation.However, during the procedure, it was noted that the balloon shaft was kinked.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was good.However, returned device analysis revealed balloon torn longitudinally.
|
| |
|