| Model Number |
ZCT525 |
| Medical Device Problem Code |
Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/28/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted; give date: not applicable as the lens was inserted and removed.If explanted; give date: not applicable as the lens was inserted and removed.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a scratch was noticed on an intraocular lens (iol) after it was inserted into the right eye of a female patient.The surgeon commented that the defect was small but very central on the optic; it looked like a small divot on the posterior surface of the iol.Reportedly, the surgeon removed and replaced the lens during the same surgical procedure without any issue.There was no patient injury and no incision enlargement.The patient was reported doing well one day post-op and very pleased with her vision.The lens was discarded by the customer.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the intraocular lens (iol) was not returned at the manufacturing site; therefore product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.The dfu states to examine the lens thoroughly to ensure particles have not become attached to it, and examine the lens optical surfaces for other defects.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Search Alerts/Recalls
|