ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT525
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/28/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted; give date: not applicable as the lens was inserted and removed.If explanted; give date: not applicable as the lens was inserted and removed.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a scratch was noticed on an intraocular lens (iol) after it was inserted into the right eye of a female patient.The surgeon commented that the defect was small but very central on the optic; it looked like a small divot on the posterior surface of the iol.Reportedly, the surgeon removed and replaced the lens during the same surgical procedure without any issue.There was no patient injury and no incision enlargement.The patient was reported doing well one day post-op and very pleased with her vision.The lens was discarded by the customer.No further information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the intraocular lens (iol) was not returned at the manufacturing site; therefore product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.The dfu states to examine the lens thoroughly to ensure particles have not become attached to it, and examine the lens optical surfaces for other defects.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6975431
Report Number9614546-2017-01026
Device Sequence Number16511467
Product Code HQL
UDI-Device Identifier05050474603417
UDI-Public(01)05050474603417(17)210902
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date09/02/2021
Device Model NumberZCT525
Device Catalogue NumberZCT525U120
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/29/2017
Supplement Date Received by Manufacturer11/14/2017
Initial Report FDA Received Date10/24/2017
Supplement Report FDA Received Date12/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age53 YR
-
-