| Medical Device Problem Codes |
No Display/Image (1183); Moisture Damage (1405); Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, the reporter contacted animas, alleging a display (damaged w/ moisture) issue.It was noted that the display screen was damaged and there was moisture behind the display.There was no indication that the device caused or contributed to an adverse event.This complaint is being reported because the issue may impact the user's ability to read some or all of the information on the screen which may result in over or under delivery.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up #1: date of submission (b)(6) 2017; device evaluation: the device has been returned and evaluated by product analysis on (b)(6) 2017 with the following findings: on investigation, moisture was confirmed behind the display lens.The display lens was not damaged.Leak testing confirmed moisture leak around the display lens.Moisture corrosion was also found inside the pump on the printed circuit board.Investigation duplicated the complaint.
|
| |
|