• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC UNIVERSA FIRM URETERAL STENT SET; FAD STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC UNIVERSA FIRM URETERAL STENT SET; FAD STENT, URETERAL Back to Search Results
Model Number G49913
Medical Device Problem Code Split (2537)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/13/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The event is currently under investigation.A follow-up report will be submitted upon receipt of additional information or completion of the investigation.
 
Event or Problem Description
The area representative reported during an antegrade placement of a ureteral stent, the surgeon used the universa firm ureteral stent set.When the surgeon removed the tether, the pigtail was split.A new universa stent was opened and used to complete the procedure.As reported, this occurred in consecutive cases.There was no part of the device left inside the patient¿s body.No additional procedures were required due to this occurrence.There were no adverse effects or consequences to the patient.Manufacturer report #¿s 1820334-2017-03616 (first case) and 1820334-2017-03617 (second case) have been submitted to capture the consecutive cases in which the pigtail was split when the tether was removed.The surgeon expressed that he intensely dislikes the braided tethers that come with the universa stents and would prefer they come with the monofilament tethers that are sold in the united states.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: the universa firm ureteral stent was not returned for evaluation and no photographs were provided.Without the complaint device, a physical investigation was not able to be completed.The investigation included a review of the device history record, complaint history, the instructions for use, manufacturing instructions and quality control data.Based on the investigation evaluation, there is no indication that a design or process related failure mode contributed to this event.A review of production and quality documentation did not identify any specific issues with current manufacturing or quality controls that may have contributed to this incident.A review of the device history record showed there were no non-conformances identified during the manufacturing process.A review of complaint history for this product/lot number combination revealed this is the only complaint that has been received against this complaint device lot number; 7931750.Based on the provided information the definitive root cause is inconclusive.Per the quality engineering risk assessment, no further action is required.Cook medical has notified the appropriate personnel and will continue to monitor this device via the complaints database for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSA FIRM URETERAL STENT SET
Common Device Name
FAD STENT, URETERAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key6976254
Report Number1820334-2017-03616
Device Sequence Number19115387
Product Code FAD
UDI-Device Identifier10827002499130
UDI-Public(01)10827002499130(17)200522(10)7931750
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K961446
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG49913
Device Catalogue NumberUFH-724-R
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/13/2017
Supplement Date Received by Manufacturer12/07/2017
Initial Report FDA Received Date10/25/2017
Supplement Report FDA Received Date12/13/2017
Patient Sequence Number1
-
-