| Model Number |
7205357 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/04/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the hand piece of the mini motor did not work during the surgery.The pedal was activated and the console looked as if it was working but the blade did not move, the pedal was changed, the blade was changed, but it did not work.After trying several times and disconnecting and reconnecting the console, it started to whistle and an engine error sign came out.Doctor then decided to perform an open surgery to complete the procedure.No patient injuries were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Mdu failed for motor stall error and overheating.Cause of motor stall and overheating is a jammed motor/gearbox.The motor/gearbox assembly could not be removed from the housing for further assessment because this type of mdu has the motor/gearbox potted inside of housing.The complaint was confirmed and the root cause has been determined to be a jammed motor/gearbox assembly.Factors which can contribute to jammed motor/gearbox include cleaning and sterilization methods and the chemicals involved.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
|
| |
|