• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. APOLLO WAND; GWG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. APOLLO WAND; GWG Back to Search Results
Catalog Number AP9
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/27/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the apollo wand (wand) leaked at the connection between the apollo generator (generator) and the acrylonitrile butadiene styrene (abs) knob.Conclusions: evaluation of the returned device confirmed that the wand leaked.The wand was connected to an apollo system and was able to aspirate, irrigate, and vibrate; however, the wand leaked where the abs knob connects to the generator.The root cause of the leak could not be determined.Penumbra wands are visually inspected during in-process inspection and during quality inspection after manufacturing the manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a microneurosurgery procedure, the physician noticed the apollo wand (wand) was leaking from the connection to the apollo generator.The damaged wand was noticed prior to use in the patient and therefore, was not used in the procedure.The procedure was completed using a new wand.
 
Additional Manufacturer Narrative
Please note that section box 4.Device manufacture date was incorrectly reported on the initial mfr report and is being corrected on this follow-up # 1 mfr report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APOLLO WAND
Common Device Name
GWG
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key6976979
Report Number3005168196-2017-01859
Device Sequence Number13820430
Product Code GWG
UDI-Device Identifier00814548016344
UDI-Public00814548016344
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K152699
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date04/20/2018
Device Catalogue NumberAP9
Device Lot NumberC02159
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2017
Initial Date Received by Manufacturer 09/27/2017
Supplement Date Received by Manufacturer09/27/2017
Initial Report FDA Received Date10/25/2017
Supplement Report FDA Received Date03/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-