Results: the apollo wand (wand) leaked at the connection between the apollo generator (generator) and the acrylonitrile butadiene styrene (abs) knob.Conclusions: evaluation of the returned device confirmed that the wand leaked.The wand was connected to an apollo system and was able to aspirate, irrigate, and vibrate; however, the wand leaked where the abs knob connects to the generator.The root cause of the leak could not be determined.Penumbra wands are visually inspected during in-process inspection and during quality inspection after manufacturing the manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
During preparation for a microneurosurgery procedure, the physician noticed the apollo wand (wand) was leaking from the connection to the apollo generator.The damaged wand was noticed prior to use in the patient and therefore, was not used in the procedure.The procedure was completed using a new wand.
|