| Date of Event |
09/26/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported that the cook-swartz doppler probe doppler wires were coming loose post operatively, which necessitated the reopening of the patient's neck incision to determine what resulted in the loss of the doppler signal.The doppler malfunction reportedly directly contributed to the additional medical intervention.The patient's original procedure was a pharyngectomy with anterior lateral thigh free tissue transfer.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: complaint was confirmed via customer testimony.No physical evaluation could be performed as device was discarded by the customer.The device history record was reviewed, as well as manufacturing and quality control record.There is no evidence that defective product was shipped.The drawing of the finished assembly was also reviewed and there are no signs that the cuff was detached prior to shipment or that the device was not manufactured to current specifications.Investigation results concluded that the reported event is within the scope of a completed internal corrective action plan.A summary of the investigation has been sent to the complainant.
|
| |
|