ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ALLURE QUADRA RF CRT-P IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3242
Device Problems Pacemaker Found in Back-Up Mode (1440); Operating System Version or Upgrade Problem (2997)
Patient Problem Muscle Stimulation (1412)
Event Date 09/25/2017
Event Type  Injury  
Event Description
It was reported that the patient presented in-clinic for follow-up. During firmware upgrade, the pulse generator exhibited backup vvi mode. The patient became symptomatic and experienced pocket stimulation. The device was successfully restored through a device download. After reprogramming the device, the patient's symptoms subsided. The patient elected not to upgrade the firmware.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameALLURE QUADRA RF CRT-P
Type of DeviceIMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key6977533
MDR Text Key90246729
Report Number2017865-2017-33600
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial
Report Date 10/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date03/31/2017
Device Model NumberPM3242
Device Lot NumberA000011749
Other Device ID Number05414734507912
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/25/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/18/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 10/25/2017 Patient Sequence Number: 1
-
-