• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM Back to Search Results
Model Number MT22608-BLU
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/03/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Dexcom was made aware on (b)(4) 2017 that on (b)(6) 2017 the receiver had no audio output.No additional event or patient information is available.The receiver was returned for evaluation.An exterior visual inspection was performed and passed.The receiver was charged and will boot.A manual test was performed and failed as no sound was heard from the speaker.The reported event of no audio output was confirmed.The root cause was determined to be a defective speaker.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
MDR Report Key6977770
Report Number3004753838-2017-91693
Device Sequence Number13735823
Product Code MDS
UDI-Device Identifier20386270000204
UDI-Public20386270000204
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMT22608-BLU
Device Catalogue NumberSTK-CR-BLU
Device Lot Number5202420
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2017
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2017
Initial Report FDA Received Date10/25/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age12 YR
-
-