Model Number 7577045
Medical Device Problem Codes Contamination (1120); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Patient Problem/Medical Problem (2688)
Date of Event 08/25/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No radiographs, photograph or product were provided to confirm the reported event.No root cause can be determined at this time, however due to the length of time from surgery to infection as well as conveyed culture results suggest that the infection was community acquired.Labeling review:.All non-sterile parts should be cleaned and sterilized before use." ".Potential adverse events and complications: as with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative and postoperative complications known to occur include: early or late infection which may result in the need for additional surgeries." not returned for evaluation.
 
Event or Problem Description
On (b)(6) 2017 a patient underwent an interlaminar lumbar instrumented fusion procedure at l5-s1 levels without issue.During follow up it was observed that the incision site was not healing.On (b)(6) 2017 a revision procedure took place to remove the implants without reported issue.
 
Additional Manufacturer Narrative
Supplemental reported submitted to update based on information received.Culture results returned negative.No infection was detected, no product malfunction.Patient is reported to be doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPINOUS PROCESS PLATE SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
NUVASIVE, INC.
7475 lusk blvd
san diego CA 92121
MDR Report Key6977868
Report Number2031966-2017-00151
Device Sequence Number1272828
Product Code KWP
UDI-Device Identifier00887517062352
UDI-Public00887517062352
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K073278
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Model Number7577045
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/29/2017
Supplement Date Received by Manufacturer01/08/2018
Initial Report FDA Received Date10/25/2017
Supplement Report FDA Received Date01/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
-
-