| Medical Device Problem Codes |
Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6), the reporter contacted animas, alleging a call service alarm (call service alarm issue) issue.It was reported that there was a 069 call service alarm.There was no indication that this issue lead to an adverse event.This is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up #1: the device has been returned and evaluated by product analysis on 11-nov-2017 with the following findings: a review of the pump¿s black box and alarm history showed a call service 069 alarm, which is written in the pump's black box as cs87.During testing, a hard call service 69 alarm was reproduced at the rewind step.The cs 069 failure is defined as a language file corruption while retrieving the language text.The language corruption issue causing the cs69 alarm was found to be due to a malfunction in the u39 flash chip, a component on the printed circuit board.
|
| |
|