Model Number H7493926220300 |
Device Problems
Bent (1059); Failure to Advance (2524)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/11/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device is a combination product.(b)(4).A synergy ous mr 3.00 x 20mm stent delivery system (sds) was returned for analysis.A visual examination of the crimped stent found distal stent damage.The most distal stent struts row was damaged; struts were lifted from the stent profile.The crimped stent od (outer diameter) of the undamaged portion of the stent was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of shaft polymer extrusion revealed no issues.A visual and tactile examination found damage to the tip.No other issues were identified during the product analysis.The most probable root cause is operational context as device performance was limited due to anatomical/procedural factors.(b)(4).
|
|
Event Description
|
Reportable based on device analysis completed on 08-oct-2017.It was reported that crossing difficulties were encountered and the distal tip got kinked.The target lesion was located in the severely tortuous and moderately calcified left circumflex obtuse marginal branch.A 3.00 x 20 synergy¿ drug-eluting stent was advanced but failed to cross the lesion.The device was removed and the distal tip was found to be kinked.The procedure was completed with another of the same device.No patient complications were reported.However, returned device analysis revealed distal stent damage.
|
|
Search Alerts/Recalls
|