• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD BHR DYSPLASIA CUP 56MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD BHR DYSPLASIA CUP 56MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74122256
Health Effect - Clinical Codes Patient Problem/Medical Problem (2688); No Code Available (3191)
Date of Event 09/19/2017
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that revision surgery was performed.
 
Additional Manufacturer Narrative
It was reported that right hip revision surgery was performed due to an infection.During the first stage of the revision the modular head, bhr dysplasia cup, 2 x dysplasia screws, 1 x cortical screw, sleeve, stem and centraliser were removed.A spacer was implanted before completion of the revision 3 days later.As of today, device return and additional information has been requested for this complaint but has not become available.In the absence of the actual devices, the production and sterilisation records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.The available medical documents were reviewed.Based on the provided operation record from a two-stage revision, the reason for this revision can be assumed to be due to infection.Further investigation was not provided in this regard.An infection developed more than two years after implantation is a late infection.Late infections are normally acquired hematogenously and are not the result of intraoperative contaminations.Since no results of a microbiological analysis were received, further assessment regarding these circumstances could not be performed.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint.The available information suggests that a late infection has occurred, necessitating the total revision of the components which is not related to the devices originally implanted or their sterility and is not a device related revision.However if the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BHR DYSPLASIA CUP 56MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key6979016
Report Number3005975929-2017-00383
Device Sequence Number13946838
Product Code NXT
UDI-Device Identifier03596010552372
UDI-Public03596010552372
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P040033
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2017
Reporter Type Manufacturer
Report Source consumer,foreign
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 10/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2012
Device Catalogue Number74122256
Device Lot Number07EW11849
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/17/2017
Supplement Date Received by Manufacturer10/17/2017
Initial Report FDA Received Date10/26/2017
Supplement Report FDA Received Date07/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DYSPLASIA SCREW, # 74500072, LOT # 07AT05161; DYSPLASIA SCREW, # 74500072, LOT # 0EM11017; FEMORAL CENTRALISER, # 71313104, LOT # 09AM16182; FEMORAL STEM, # 71312102, LOT # 09FT32521; FIXATION SCREW, # 3571149754, LOT # 05CT88059; MODULAR HEAD, # 74222148, LOT # 08LW20316; MODULAR SLEEVE, # 74222100, LOT # 08LW19877
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age58 YR
-
-