| Medical Device Problem Codes |
Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017 the reporter contacted animas and reported that call service alarm had been emitted by the pump more often than expected by the user.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 29-nov-2017 with the following findings:during investigation, the pump was powered on and a hard call service 069 alarm was shown that could not be reset.A review of the pump¿s black box and alarm history showed a call service 069 alarm, which is written in the pump's black box as cs87.The cs 069 failure is defined as a language file corruption while retrieving the language text.The language corruption issue causing the cs69 alarm was found to be due to a malfunction in the u39 flash chip, a component on the printed circuit board.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
| |
|