• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. TURON SHOULDER; TURON PEGGED GLENOID, SZ 42MM, EPLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. TURON SHOULDER; TURON PEGGED GLENOID, SZ 42MM, EPLUS Back to Search Results
Model Number 520-01-006
Health Effect - Clinical Code No Code Available (3191)
Date of Event 09/26/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - revision to a reverse shoulder because of patient rotator cuff tear.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery is due to a tear in the patient's rotator cuff.The in-vivo length of patient service for the implant was 2.4 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions, design criteria and specifications in effect at the time the part was manufactured were met.The complaint investigation history was reviewed and no trends or on-going issues were deemed to be present or in need of review.This event is deemed as non-product related.The root cause for this event was the patient revision had a rotator cuff tear and required a revision.The scope of this investigation is limited without having the explanted parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required since there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
TURON PEGGED GLENOID, SZ 42MM, EPLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key6980003
Report Number1644408-2017-00935
Device Sequence Number444198
Product Code KWS
UDI-Device Identifier00888912144896
UDI-Public(01)00888912144896
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K123982
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2019
Device Model Number520-01-006
Device Catalogue Number521-01-242
Device Lot Number799N1019
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/26/2017
Supplement Date Received by Manufacturer11/06/2017
Initial Report FDA Received Date10/26/2017
Supplement Report FDA Received Date11/20/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
(B)(4),LOT 799N1019; (B)(4),LOT 878C1404; (B)(4),LOT 911C1036
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
-
-