| Model Number |
M004218500 |
| Medical Device Problem Code |
Device Remains Activated (1525)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/06/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the footswitch cannot deactivate rf delivery.The target lesion was located in the floor of the right atrium.A maestro foot switch and a maestro 4000¿ controller were selected for use.During the procedure, a successful typical flutter cti ablation using a maestro 4000 and a foot pedal, along with a mifi nav xp and a rhythmia system was performed.After approximately 30 lesion, physician stepped off the device.However, rf energy continued to deliver.It was stopped by shutting down the generator remotely.No patient complications reported.
|
| |
|
Search Alerts/Recalls
|