Model Number 186715650
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/06/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is available for evaluation.Investigation will be conducted.Follow up will be filed with findings.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The screw head was loose from the screw body.The screw has never been used before.
 
Additional Manufacturer Narrative
Device was not returned for evaluation.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.A trend analysis was conducted.No emerging trends were found requiring further actions.Without the returned of the device we are unable to confirm the reported issue or identify the root cause.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.Should more information and/or the sample be provided at a later time, this complaint will be reopened and device evaluated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device not returned to manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
DEPUY SYNTHES SPINE
325 paramount drive
raynham MA 02767
MDR Report Key6980968
Report Number1526439-2017-10921
Device Sequence Number118898
Product Code NKB
UDI-Device Identifier10705034122366
UDI-Public(01)10705034122366
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K160904
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model Number186715650
Device Catalogue Number186715650
Device Lot Number126942
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/09/2017
Supplement Date Received by Manufacturer11/13/2017
Initial Report FDA Received Date10/26/2017
Supplement Report FDA Received Date11/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-