| Model Number |
ZCB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Edema (1820); No Code Available (3191)
|
| Date of Event |
09/06/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As the lens remains implanted.Concomitant medical products: 1mtec30 cartridge and dk7796 handpiece serial and lot number unknown.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zcb00 26.0 diopter intraocular lens (iol) was implanted in the left eye (os) on (b)(6) 2017 with use of a 1mtec30 cartridge.Post-operatively, one day after surgery, the doctor diagnosed the patient with hypopyon, fibrin membrane, snowstorm, which cleared quickly on intense steroids.In addition to steroids, the patient also received omidria.The doctor also stated the patient had significant endothelial folds-edema, possible toxic anterior syndrome (tass), and fibrin.The case was routine - nothing out of ordinary, this case seemed more c/w tass like presentation.No iol explant has been or will be performed.No additional information was provided.This report represents the intraocular lens and a separate report will be provided for the cartridge.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the product was not returned to the manufacturing site as it remains implanted.Therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints have been received for this production order number.Sterility records review: the sterilization records for this lot were reviewed and no deviations were found.The product was processed according to requirements and met specifications.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Search Alerts/Recalls
|