| Model Number |
150D2 |
| Medical Device Problem Codes |
Detachment Of Device Component (1104); Device Dislodged or Dislocated (2923); Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/28/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event has been treated in leventon (b)(4) as an incidence ((b)(4)).According to the results obtained, the defect reported by the client is due to a random human error during the manual assembly of the pump.
|
| |
|
Event or Problem Description
|
|
Defective dosi-fuser 150d2 pump.Prior to filling the pump, technician noted that "level indicator" has dislodged from the top of the balloon and was floating freely within the container.This could pose a problem in determining if a patient's infusion has completed.
|
| |
|