• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX INC. IBALANCE TKA FEMORAL IMPLANT PS,CEMENTED SZ 4 RGHT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX INC. IBALANCE TKA FEMORAL IMPLANT PS,CEMENTED SZ 4 RGHT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number AR-516-4R
Health Effect - Clinical Codes Bacterial Infection (1735); Fall (1848)
Date of Event 07/12/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No further patient information was provided at the time of this report or made available in response to arthrex's follow-up requests.The device was not returned for evaluation and the contribution of the device to the reported event could not be determined.The information received regarding the event is not sufficient to determine the cause of the event.A dhr review for this catalog number, ar-516-4r and lot number 108761406 combination found no related information.
 
Event or Problem Description
On 9/12/2017 it was reported via the patient's attorney that a patient had undergone a revision tka procedure.It was reported that patient's first issues with the explanted device were instability, pain and motion which arose in (b)(6) 2015.At the time of initial report the following arthrex product was indicated: ar-503-ttte, lot 1206173.Neither the original surgery date or revision surgery date were provided during initial report.(b)(6).Additional information received on 10/02/2017: on (b)(6) 2014 patient underwent a bilateral tka procedure.Patient has experienced constant pain since original surgery.On (b)(6) 2016 exam and radiographs were performed.On (b)(6) 2016 surgeon performed a revision right tka due to pain and tibial loosening.During the revision procedure the surgeon explanted the ibalance tka tibial tray, ar-503-ttte, lot 1206173 ((b)(4)); ibalance patella implant dome, ar-504-psc9, lot 113601411 ((b)(4)); ibalance tka bearing, ar-513-be13, lot 113601345 ((b)(4)); and ibalance tka femoral implant, ar-516-4r, lot 108761406 ((b)(4)).To complete the procedure the surgeon used another manufacturer's product.Patient was a female.Additional information received on 10/9/2017: medical records dated (b)(6) 2015 notes "that the patient fell off a ladder on (b)(6) 2015 and fractured her right ankle.Patient underwent surgery to repair her right ankle on (b)(6) 2015.Medical records (b)(6) 2016 note a (b)(6) result".Medical records also noted patient had a bmi of approx.31 which is contraindicated for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IBALANCE TKA FEMORAL IMPLANT PS,CEMENTED SZ 4 RGHT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
ARTHREX INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key6983250
Report Number1220246-2017-00403
Device Sequence Number19003957
Product Code JWH
UDI-Device Identifier00888867109339
UDI-Public00888867109339
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K081127
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2016
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Attorney
Type of Report Initial
Report Date (Section B) 10/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/01/2019
Device Catalogue NumberAR-516-4R
Device Lot Number108761406
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2017
Initial Report FDA Received Date10/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age58 YR
-
-