DEPUY SYNTHES POWER TOOLS BATT-HANDPIECE MOD F/TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 05.001.201 |
Device Problem
Difficult To Position (1467)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/11/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the battery handpiece device no longer held the cutter device.It was not reported if there was a delay to the surgical procedure or if a spare device was available for use.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device bearings did not function and were defective.The device also failed pretest for check of free moving and check for leakage.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|