| Catalog Number |
71675511 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
10/11/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a delay occurred during the surgery because several guide pins were bent and had problems attaching to the nail.
|
| |
|
Additional Manufacturer Narrative
|
|
It was noted that the hospital attributed this event to an user error.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that when inserting the anti rotation bar into the femoral head the operators noticed it converged with the 3.2mm x 343mm guide pin.They tried a new guide pin as they discovered it was bent, only to have the same result.When they removed the new one it was bent too.They removed the whole nail and noticed the set screw had dropped possibly bending the guide pin on entry.They were not sure why the set screw had dropped but there was some feeling the 4.7mm hex had been incorrectly used to try to attach the nail to the jig.Surgical delay occured due to this but to no adverse outcome to the patient.
|
| |
|