| Model Number |
E SERIES |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device would intermittently not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to zoll medical corporation.Instead, the suspect system board was returned.The reported malfunction was observed and attributed to the lithium battery on the system board.The customer received a replacement system board.Zoll medical corporation recommends replacement of the lithium battery on the system board every 5 years.This device is approximately 8 years old from date of manufacture.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|