Device Problem Crack (1135)
Patient Problem No Patient Involvement (2645)
Event Date 10/02/2017
Event Type  malfunction  
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The bellows canister was replaced to resolve the issue.No report of patient involvement.(b)(4).The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.
 
Event Description
The hospital reported a cracked bellows housing.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 650 A1
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi
CH 
Manufacturer Contact
john szalinski
540 w. northwest highway
barrington, IL 
MDR Report Key6987056
MDR Text Key91204587
Report Number9710602-2017-00271
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 10/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Other Device ID NumberSEE BLOCK H10
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/02/2017
Initial Date FDA Received10/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/15/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-