Model Number IS3000 A60P9B
Device Problems Electrical /Electronic Property Problem (1198); Defective Component (2292); Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/11/2017
Event Type  malfunction  
Manufacturer Narrative
Intuitive surgical inc.(isi) received the unit involved with this complaint and completed the device evaluation.Failure analysis confirmed the customer reported failure.The illuminator was installed into a test system and it failed to power on with error 297 and 48238.This complaint is being reported due to a da vinci system malfunction rendering the da vinci system unavailable for use after the start of a surgical procedure.Although no patient harm occurred, if this malfunction were to recur it could cause or contribute to an adverse event.
 
Event Description
It was reported that during a da vinci-assisted prostatectomy procedure, there was smoke and a burnt smell coming from the back of the vision side cart (vsc).Customer powered system off and it booted up with an error 297.The intuitive surgical, inc.(isi) technical support engineer (tse) reviewed the system logs and noted error 297 and 48238 which pointed to the illuminator.The isi tse advised the customer to use a third party illuminator to continue the procedure.The planned surgical procedure was completed with an alternate light source and no patient harm, adverse outcome or injury was reported.An isi field service engineer (fse) was dispatched to the facility and replaced the illuminator to resolve the issue.The illuminator is a component of the da vinci system that contains a high intensity light source to illuminate the surgical site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DA VINCI SI SURGICAL SYSTEM
Type of Device
ENDOSCOPIC INSTRUMENT CONTROL SYSTEM
Manufacturer (Section D)
INTUITIVE SURGICAL,INC.
sunnyvale CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC.,
950 kifer rd
sunnyvale CA
Manufacturer Contact
tabitha reed
950 kifer rd
sunnyvale, CA 
4085232420
MDR Report Key6988485
MDR Text Key90764788
Report Number2955842-2017-00721
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
Reporter Country CodeKR
PMA/PMN Number
K081137
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIS3000 A60P9B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received10/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-