Model Number S7 |
Device Problem
Material Frayed (1262)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/04/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.A medtronic representative inspected the navigation system onsite and confirmed the issue.The cord was replaced and the issue was resolved.A full system check-out was completed following part replacement and all tests passed.Full system functionality was confirmed and the system was returned to service.The suspect power cord has been received by the manufacturer for evaluation.However, results are not available at this time.
|
|
Event Description
|
A site representative reported that outside of a procedure, the system was observed to have a frayed power cord.There was no patient identified when this issue was identified.
|
|
Manufacturer Narrative
|
The analysis of the suspect power cable was completed by medtronic personnel.It was reported that there were internal wires exposed, however no bare conductors were exposed.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
|