Model Number 72200616S
Device Problems Loss of Power (1475); Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/29/2017
Event Type  malfunction  
Manufacturer Narrative
A visual inspection was performed and no issues were noted. Complaint of power failure could not be confirmed. Product had power but failed functional testing with hand piece error when inserted into mdu receptacle on test control unit. Cause of hand piece errors is a corroded forward button spring that was stuck in the ¿on¿ position. The complaint was confirmed and the root cause has been determined to be corrosion of the forward button assembly. Factors that could have contributed to the event include a buildup of corrosion/debris around the button assembly from cleaning chemical fluid ingression over time. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Event Description
It was reported by customer that device had no power and presented button stuck in "on" position. No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSVC REPL,MDU, HAND CNTRL, PWRMX
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key6989047
MDR Text Key91117086
Report Number1643264-2017-01258
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial
Report Date 10/30/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/30/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number72200616S
Device Catalogue Number72200616S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2017
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/17/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-