Model Number MSERIES CCT
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that the device would intermittently not power up.Complainant did not indicate that there was any patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to the lithium battery on the system board.The device operator's guide instructs users to replace the lithium battery on the device every 5 years.This device is approximately 10 years old.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M SERIES
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key6990152
Report Number1220908-2017-02702
Device Sequence Number684175
Product Code MKJ
UDI-Device Identifier00847946014421
UDI-Public00847946014421
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K011865
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberMSERIES CCT
Device Catalogue NumberM SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2017
Initial Date Received by Manufacturer 10/11/2017
Supplement Date Received by Manufacturer10/11/2017
Initial Report FDA Received Date10/31/2017
Supplement Report FDA Received Date11/16/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-