| Catalog Number |
CXI-2.6-18-90-P-NS-ANG |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/16/2015
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
|
| |
|
Event or Problem Description
|
|
The cxi was used with a cook micropuncture device and a wire guide from another manufacturer.When the physician tried to take out the wire, it was not possible even with much force.The cxi tore and the physician had to take out all of the products together.It was noted that all 3 products were melted together.However, nothing was lost in the patient's vessel.The investigation further noted that the dilator was melted onto the cxi.There were no reported adverse effects to the patient or additional procedures as a result of this product problem.
|
| |
|