Model Number AT75184
Medical Device Problem Codes Material Frayed (1262); Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/01/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was returned for evaluation.A visual inspection found frayed fibers on the barrel of the balloon.An inflation attempt was made and a pinhole rupture was noted in the balloon.Therefore, the investigation was confirmed for a pinhole rupture and was confirmed for frayed fibers.Per the reported event details, the balloon ruptured within a stent.It is possible that an interaction between the balloon and stent caused or contributed to the identified failures.However, the definitive root cause could not be determined based upon the available information.Labeling review: the current ifu (instructions for use) states: warnings: do not exceed the rbp recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended.Precautions: if resistance is felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and guidewire/introducer sheath as a single unit.Use of the atlas pta dilatation catheter: apply negative pressure to fully evacuate fluid from the balloon.Confirm that the balloon is fully deflated under fluoroscopy.While maintaining negative pressure and the position of the guidewire, withdraw the deflated dilatation catheter over the wire through the introducer sheath.Use of a gentle counterclockwise motion may be used to help facilitate catheter removal through the introducer sheath.(b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during the first inflation, the balloon catheter was inflated to 1-2 atm past nominal pressure, post dilating a stent in the iliac, when the balloon allegedly ruptured.It was further reported by the health care provider that the balloon allegedly had a pinhole rupture.There were no reported retraction difficulties through the stent or the sheath.Reportedly, the balloon catheter was exchanged over the guidewire for another and the procedure was completed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS PTA BALLOON DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6990880
Report Number2020394-2017-01471
Device Sequence Number1227900
Product Code DQY
UDI-Device Identifier00801741062490
UDI-Public(01)00801741062490
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K120971
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/28/2019
Device Model NumberAT75184
Device Catalogue NumberAT75184
Device Lot NumberGFAY3060
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2017
Initial Report FDA Received Date10/31/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-