ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Model Number FEM07080
Medical Device Problem Codes Positioning Failure (1158); Fracture (1260); Misfire (2532)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/02/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or no medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during a stent graft placement procedure intended for the basilic vein, the stent allegedly deployed partially and did not deploy any further.Reportedly, the outer sheath of the device fractured.Another device was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
A manufacturing review was performed.The lot records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.The review shows that no remarkable incidents occurred during the manufacturing process.No relevant manufacturing process changes were implemented, that could have led to the event reported.No additional complaint has been reported for this lot number previously.The stent graft delivery system was returned for evaluation.Based on the condition of the sample returned the reported issue could be confirmed.The stent graft was found partially deployed and the outer sheath was found elongated and fractured, which indicates that increased friction forces did occur during the attempt to deploy the stent graft.Potential factors which may have caused or contributed to the reported issue have been considered.However, a definite root cause for the reported event could not be determined.In reviewing the labeling supplied with this product it was found that the instructions for use (ifu) sufficiently address the potential risks.The ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device." regarding preparation of the device the ifu states that "prior to loading the endovascular system over a guide wire, both ports must be flushed with sterile saline.Flushing these lumens will also facilitate stent graft deployment." regarding the anatomy of the placement site the ifu states: "the safety and effectiveness of the device when placed across a tight bend has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure.".
 
Event or Problem Description
It was reported that during a stent graft placement procedure intended for the basilic vein, the stent allegedly partially deployed and the outer sheath of the device allegedly fractured.Reportedly, another device was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM  76227
MDR Report Key6991035
Report Number2020394-2017-01473
Device Sequence Number444350
Product Code PFV
UDI-Device Identifier04049519008455
UDI-Public(01)04049519008455
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P130029
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/13/2020
Device Model NumberFEM07080
Device Catalogue NumberFEM07080
Device Lot NumberANBR1723
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2017
Initial Date Received by Manufacturer 10/04/2017
Supplement Date Received by Manufacturer01/18/2018
Initial Report FDA Received Date10/31/2017
Supplement Report FDA Received Date01/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-