| Catalog Number |
309.450 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
10/05/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device is an instrument and is not implanted/explanted.A device history record (dhr) review was performed for part # 309.450, lot # 8888465: manufacturing location: (b)(4) manufacturing date: 28.Apr.2014: no non-conformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.The device has been received and the product evaluation is in progress.No conclusion can be drawn.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that initially a patient underwent a placement of a medial malleolar screw for an ankle valgus on an unknown date to correct a deformity.A hardware removal was scheduled on (b)(6) 2017 due to the deformity healing.During the removal, a 4.0mm cannulated screw stripped.As the surgeon was removing the screw from the medial malleous, the screw head stripped.The surgeon used pliers to back the screw out.The head then broke off, so the surgeon used a trephine to open around the shaft of the screw.The trephine then broke off inside the patient.The surgical team then used a conical extractor to remove the screw shaft.The screw then broke off inside the patient.Using the hardware removal tools to extract the screw, the surgery was able to be completed with the removal of the entire screw.There was about a ten (10) minute surgical delay.The patient status is good.Concomitant devices reported: pliers (part # unknown, lot # unknown, qty 1), conical extractor (part # unknown, lot # unknown, qty 1); hardware removal tools (part # and lot # unknown, quantity unknown).This report is for one (1) hollow reamer-complete for 4.5mm screws.This is report 3 of 4 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Product development investigation was completed on the hollow reamer (309.450).The investigation summary indicates that: this complaint involves 4 parts.Customer quality (cq) engineering investigation: complaint condition of ¿broken ¿ intraoperatively¿ was confirmed for the two out of four devices which are unknown cannulated screw and the reamer tube p/n 309.480.Conical screw extractor p/n 309.039 and the hollow reamer (p/n 309.450) were found to be intact with no broken or missing fragments during cq investigation.The replication of the complaint condition at customer quality (cq) is not applicable for the broken devices as the devices were visibly confirmed to be broken.The fragment of the screw was received stuck inside the conical extractor (p/n 309.039) and the hollow reamer (p/n 309.450) was found stuck with the reamer tube p/n 309.480 on the coupling end.The two devices could not be separated using pliers at the cq location.The overall complaint is confirmed.Visual inspection, device history record (dhr) review, dcrm review and drawing review were performed as part of this investigation.A possible collision of the reamer tube (p/n 309.480) with an implant may have caused an impact that lead to the device being stuck (cold welded) to the hollow reamer (309.450) at the proximal end where they are threaded together.It is also possible that the user was unaware of the reverse threading on the hollow reamer (p/n 309.450) which requires clockwise rotation to loosen the devices.And the two devices got stuck together during an attempt to separate them at the hospital following the surgery.The exact root cause could not be determined.Device used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|