| Model Number |
U29525 |
| Medical Device Problem Code |
Split (2537)
|
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The unit was returned with no complaint listed.Upon triage, the following issue was found: damaged power cord with exposed copper.
|
| |
|
Additional Manufacturer Narrative
|
|
The scd 700 was evaluated and the review showed that the unit had a damaged power cord, with copper wire was exposed.From the area in which the damage occurred it is possible this could be related to wear over time from the wrapping procedure of the cord around the bed hook.The cause of the reported condition for the damaged power cord was due to wear from potential wrapping technique overtime.A review of the device history record shows that this unit was manufactured and was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|