• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9500-44
Device Problem Component Missing (2306)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/10/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(6) 2017, that on (b)(6) 2017, a missing sensor wire occurred.The sensor was inserted into the abdomen on (b)(6) 2017.No medical intervention was reported.Additional event or patient information is not available.No product was provided for evaluation.The complaint confirmation was unable to be determined.A root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key6993948
MDR Text Key91311549
Report Number3004753838-2017-98260
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 10/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/02/2018
Device Model Number9500-44
Device Catalogue NumberSTS-GF-001
Device Lot Number5223522
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/10/2017
Initial Date FDA Received11/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-