| Model Number |
SD300LG |
| Medical Device Problem Codes |
Detachment Of Device Component (1104); Unstable (1667)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/07/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation of the locatable guide found that the tip was not secured during construction of the device and the crimp length was not adequate for proper crimping as crimped in manufacturing.An internal corrective action has been initiated.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The locatable guide was returned with another device for analysis and the locatable guide was found with the sensor tip detached from the device.There was no report of patient injury, death, or other serious adverse event.
|
| |
|