| Model Number |
T505U225 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Non specific EKG/ECG Changes (1817); High Blood Pressure/ Hypertension (1908); Occlusion (1984)
|
| Date of Event |
09/26/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during the implant of this 25 mm bioprosthetic valve, the valve was explanted and replaced with a 23 mm bioprosthetic valve of a different model.The blood pressure was elevated and electrocardiogram (ecg) showed elevated st segments.No failure mechanism of the 25 mm bioprosthetic valve was reported.The surgeon reported the 25 mm valve occluded the right coronary ostia and the same effective orifice area (eoa) was accomplished with the 23 mm valve.Following the implant of the 23 mm bioprosthetic valve, no further adverse patient effects were reported.
|
| |
|