COVIDIEN MFG SOLUTIONS S.A. MAHURKAR; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number 8813793009
Device Problems Occlusion Within Device (1423); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/09/2017
Event Type  malfunction  
Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.If the sample is received or if additional information pertinent to the incident is obtained a follow-up report will be submitted.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, on (b)(6) 2017 during insertion of a new catheter the blue lumen was occluded and the guide wire could not pass.As a result, the catheter was not used.There was no patient injury.Anecdotally the customer stated this happened on another patient but they had no specific information regarding that event.
 
Manufacturer Narrative
Evaluation summary: a device history (dhr) review revealed no discrepancies that may have contributed to a complaint of this failure mode.All quality assurance testing performed during manufacturing was acceptable.The quality assurance review of the visual, physical and dimensional evaluation results indicated that the product met specification requirements.In addition, all dhr are reviewed for accuracy prior to product release.The actual sample involved in the reported incident was not returned for evaluation.No additional information, pictures or videos were received.Consequently, it was not possible to evaluate it as part of a comprehensive failure investigation.No probable cause was found since no sample, picture or video were received for testing, therefore the defect is not confirmed.If the sample is returned in the future, this complaint will be re-opened for further investigation.No trends or triggers have been found.No action is not deemed necessary at this time.It must be noted that in-process controls are in place to prevent nonconforming product from leaving the manufacturing operations.This complaint will be used for tracking and trending purposes.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAHURKAR
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
Manufacturer (Section G)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
Manufacturer Contact
lisa matheuw
15 hampshire street
mansfield, MA 02048
2034925563
MDR Report Key6997542
MDR Text Key91094656
Report Number3009211636-2017-05340
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier20884521004808
UDI-Public20884521004808
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K943349
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/28/2021
Device Model Number8813793009
Device Catalogue Number8813793009
Device Lot Number1702600150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/11/2017
Initial Date FDA Received11/02/2017
Supplement Dates Manufacturer Received07/16/2018
08/08/2018
Supplement Dates FDA Received08/06/2018
10/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/28/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
-
-