COVIDIEN MFG SOLUTIONS S.A. MAHURKAR CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
|
Back to Search Results |
|
Model Number 8813793009 |
Device Problems
Occlusion Within Device (1423); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/09/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation summary: a device history (dhr) review revealed no discrepancies that may have contributed to a complaint of this failure mode.
All quality assurance testing performed during manufacturing was acceptable.
The quality assurance review of the visual, physical and dimensional evaluation results indicated that the product met specification requirements.
In addition, all dhr are reviewed for accuracy prior to product release.
The actual sample involved in the reported incident was not returned for evaluation.
No additional information, pictures or videos were received.
Consequently, it was not possible to evaluate it as part of a comprehensive failure investigation.
No probable cause was found since no sample, picture or video were received for testing, therefore the defect is not confirmed.
If the sample is returned in the future, this complaint will be re-opened for further investigation.
No trends or triggers have been found.
No action is not deemed necessary at this time.
It must be noted that in-process controls are in place to prevent nonconforming product from leaving the manufacturing operations.
This complaint will be used for tracking and trending purposes.
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
The incident sample has been requested but to date has not been received for evaluation.
If the sample is received or if additional information pertinent to the incident is obtained a follow-up report will be submitted.
A good faith effort will be made to obtain the applicable information relevant to the report.
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, on (b)(6) 2017 during insertion of a new catheter the blue lumen was occluded and the guide wire could not pass.
As a result, the catheter was not used.
There was no patient injury.
Anecdotally the customer stated this happened on another patient but they had no specific information regarding that event.
|
|
Search Alerts/Recalls
|
|
|