| Model Number |
CS08657-02 |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that the unit did not generate an audible alarm.It was reported that there was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
One sample was received for analysis and the reported issue was confirmed.Customer speaker was removed and reconnected.Audio was found to be intermittent based on the position of the white speaker wire connection at the speaker side.Visual inspection under magnification found a cracked solder joint.Speaker was removed and retained.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that the unit did not generate an audible alarm.It was reported that there was no patient involvement.
|
| |
|