• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX (TRIVASCULAR INC.) OVATION IX; MAIN BODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX (TRIVASCULAR INC.) OVATION IX; MAIN BODY Back to Search Results
Model Number TV-AB2080-J
Medical Device Problem Codes Failure To Adhere Or Bond (1031); Leak/Splash (1354)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 10/05/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Endologix will continue to investigate the reported event.The patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.A final report will be submitted once the investigation of the reported event has concluded.
 
Event or Problem Description
An ovation ix abdominal stent graft system was implanted to treat an abdominal aortic aneurysm.The index procedure went as expected.Patient returned approximately 2 months later with an occluded left limb and a late type ia endoleak.Physician elected to re-intervene and place icast stents into the occluded left limb to open the vessel.Coils and a palmaz stent were also placed at the level of the sealing rings, which successfully resolved the occlusion and the type ia endoleak.As of the date of this report, there have been no additional patient sequelae reported.
 
Additional Manufacturer Narrative
Based on the clinical assessment for this event, the most likely cause of the loss of seal was determined to be user related, due to the off label neck and vent procedure.The significant juxtarenal angulation at the level of the renal arteries likely contributed to the event.The final patient disposition is unknown and there have been no additional patient sequelae reported.A review of the device quality records shows that the device demonstrated compliance to established procedures and specifications at the time of manufacture.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVATION IX
Common Device Name
MAIN BODY
Manufacturer (Section D)
ENDOLOGIX (TRIVASCULAR INC.)
3910 brickway blvd.
santa rosa CA 95403
MDR Report Key6998063
Report Number3008011247-2017-00071
Device Sequence Number1354082
Product Code MIH
UDI-Device IdentifierM701TVAB2080J1
UDI-Public+M701TVAB2080J1/$$3190119FS121415363
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P120006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/19/2019
Device Model NumberTV-AB2080-J
Device Catalogue NumberTV-AB2080-J
Device Lot NumberFS121415-36
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/05/2017
Supplement Date Received by Manufacturer10/05/2017
Initial Report FDA Received Date11/02/2017
Supplement Report FDA Received Date01/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
-
-