| Model Number |
92210 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/09/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation conclusion: the customer's observation was not replicated in-house with retention products.Retention products were tested with qc cutoff hcg concentrations standards.All devices showed positive results at read time and met qc specifications.No false negative results were obtained during in-house testing.Manufacturing batch record review did not uncover any abnormalities.The serum quant for the patient was reported as 94 miu/ml.A dilute urine sample may make the hcg quant level much less than the serum levels which may lead to false results.Some patients may test negative early in pregnancy if the urine hcg is below 20 miu/ml.It is recommended that if pregnancy is suspected, a first morning urine specimen should be collected 48 hours later and tested.Because the specimen associated with the complaint was not returned for investigation, a root cause could not be determined based on the information provided.Based on the information available, there is no indication of a product deficiency and no corrective action is required.
|
| |
|
Event or Problem Description
|
|
The patient's last menstrual period was reported to be (b)(6) 2017.The patient received two positive hcg results on home pregnancy tests on (b)(6) 2017.On (b)(6) 2017, the patient went to facility to confirm pregnancy.A non-first morning urine sample was collected and an alere hcg cassette from an alternate lot produced a negative hcg result at the read time of 3 minutes.However, the customer reported that a positive result developed at 4 minutes (this incident is reported under mdr 2027969-2017-00165).A serum sample was collected and produced a positive hcg quantitative result of 94 miu/ml.The patient's urine sample was then sent to a sister facility to be tested using the complaint lot of alere hcg cassettes.On (b)(6) 2017, the urine sample produced a negative hcg result at the read time of 3 minutes using the complaint lot of alere hcg cassettes.However, the customer reported that a positive result developed at 4 minutes.The sample was also tested with a siemens clinitest hcg test and produced a positive hcg result.Troubleshooting was provided to the customer.The customer was advised some patients may test negative early in the pregnancy if the urine hcg is below the 20 miu/ml cut-off, or very dilute urine specimens may not contain representative levels of hcg.Per the package insert, if pregnancy is suspected, first morning urine should be collected 48 hours later and tested or alternate confirmatory methods should be considered.
|
| |
|
Additional Manufacturer Narrative
|
|
The original investigation conclusion did not include investigation of return product.Please refer to the corrected investigation conclusion below: the customer's observation was not replicated in-house with retention and return products.Retention and return products were tested with qc cutoff hcg standards.All devices showed positive results at read time and met qc specifications.No false negative results were obtained during in-house testing.Manufacturing batch record review did not uncover any abnormalities.The serum quant for the patient was reported as 94 miu/ml.A dilute urine sample may make the hcg quant level much less than the serum levels which may lead to false results.Some patients may test negative early in pregnancy if the urine hcg is below 20 miu/ml.It is recommended that if pregnancy is suspected, a first morning urine specimen should be collected 48 hours later and tested.Because the specimen associated with the complaint was not returned for investigation, a root cause could not be determined based on the information provided.Based on the information available, there is no indication of a product deficiency and no corrective action is required.
|
| |
|
Search Alerts/Recalls
|