| Catalog Number |
H965103038031 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/05/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Although it has been indicated that the used device will be returned to angiodynamics, to date, it has not arrived.Upon evaluation of the returned device and completion of the investigation, a supplemental medwatch will be submitted.(b)(4).
|
| |
|
Event or Problem Description
|
|
As reported by hospital in (b)(6): bioflo dialysis catheter placed in patient's right internal jugular on (b)(6) 2016.It developed a hole inder the luer/hub of the catheter and was removed on (b)(6) 2017.No patient complications were reported.The used catheter will be returned to angiodynamics for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history records was performed for the indicated packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly and performance specifications.The (b)(6) 2017 angiodynamics complaint report was reviewed for the bioflo dialysis product family and the failure mode, "extension leg failure." no adverse trends were indicated.The returned catheter was visually examined and a hole was noted in the arterial extension tube adjacent to the flat section of the hub, i.E., 90 degrees from the molded parting line.There is no visible indication that the failure is related to either the manufacturing process or associated tooling.A definitive root cause was unable to be determined based on the sample examination.A capa has been initiated to provide further investigation into this and similar events, and to determine if corrective action is required.Manufacturing process controls for the dialysis catheter include visual inspection for damage or defects, 100% leak testing, and guidewire insertion testing to verify lumen patency.(b)(4).
|
| |
|
Search Alerts/Recalls
|