|
Product evaluation: 1 opened sample, part number 8229738, from lot number 0212755672 was received for analysis; there was a residue consistent with biological contaminants on the device.Based off the reported event, it is likely that any damage to the tube occurred during intubation or sometime thereafter.The cuff was inflated with 20 cc of air and it leaked out immediately which would have resulted in the reported event.When viewed under magnification, the cuff had an ¿l¿ shaped puncture in it measuring 0.04¿ x 0.05¿.There was no damage to the packaging that would indicate a cause for the observed damage.The observed defect would have been detected during an inflation test if present at the time of the test.The instructions for use indicates that damage may occur to the device during actions such as; over inflation, insertion of objects, and biting forces.The information likely indicates that the cuff was damaged during, or after intubation due to a handling issue.If information is provided in the future, a supplemental report will be issued.
|