| Catalog Number |
INT6F11C1 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Patient Problem/Medical Problem (2688)
|
| Date of Event |
09/26/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The physician intended to use an intrakit introducer device.There was no damage to packaging of the intrakit introducer.The device was removed from packaging as per ifu and inspected with no issues noted.The device was prepped per ifu with no issues.The device was properly hydrated.Resistance was encountered when the dilator was inserted into the sheath up until the last 1cm- only before the hub was connected did the resistance increase significantly.The dilator hub broke cleanly off the dilator when the physician tried to remove the dilator from the sheath while it was still in the patient's wrist.Excessive force was not used during insertion/ delivery.The device was not bent excessively prior to the detachment.No components remain in the patient.No patient injury reported.The device was replaced with a non mdt device.
|
| |
|
Additional Manufacturer Narrative
|
|
The dilator remained stuck inside the sheath once the green hub snapped off the blue dilator.The physician had to rewire through the broken dilator which was stuck inside the sheath.Once the wire was placed, the faulty sheath and dilator were removed together and a new sheath was opened and used.If information is provided in the future, a supplemental report will be issued.
|
| |
|