| Model Number |
MSERIES BI-PHASIC |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the malfunction was duplicated and the upper housing was replaced to remedy the problem.This type of damage is consistent with an overheated battery; however, the battery was not returned for evaluation.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|